Document Type : Original Research

Authors

1 Department of Medical Physics and Radiology, Faculty of Medicine, Gonabad University of Medical Sciences, Gonabad, Iran

2 Novin Institute of Medical Radiation, Shahid Beheshti University of Medical Sciences, Tehran, Iran

Abstract

Background: Polymer gel dosimeters combined with magnetic resonance imaging (MRI) can be used for dose verification of advanced radiation therapy techniques. However, the uncertainty of dose map measured by gel dosimeter should be known. The purpose of this study is to investigate the uncertainty related to calibration curve and MRI protocol for MAGIC (Methacrylic and Ascorbic acid in Gelatin Initiated by Copper) gel and finally ways of optimization MRI protocol is introduced.Materials and Methods: MAGIC gel was prepared by the Fong et al. instruction. The gels were poured into calibration vials and irradiated by 18 MV photons. 1.5 Tesla MRI was used for reading out information. Finally, uncertainty of measured dose was calculated.Results: Results show that for MAGIC polymer gel dosimeter, at low doses, the estimated uncertainty is high (≈ 18.96% for 1 Gy) but it reduces to approximately 4.17% for 10 Gy. Also, with increasing dose, the uncertainty for the measured dose decreases non-linearly. For low doses, the most significant uncertainties are σR0 (uncertainty of intercept) and σa (uncertainty of slope) for high doses. MRI protocol parameters influence signal-to-noise ratio (SNR).Conclusion: The most important source of uncertainty is uncertainty of R2. Hence, MRI protocol and parameters therein should be optimized. At low doses, the estimated uncertainty is high and reduces by increasing dose. It is suggested that in relative dosimetry, gels are irradiated by high doses in linear range of given gel dosimeter and then scaled down to the desired dose range.

Keywords

  1. Baldock C, De Deene Y, Doran S, Ibbott G, Jirasek A, Lepage M, et al. Polymer gel dosimetry. Physics in medicine and biology. 2010;55:R1. doi.org/10.1088/0031-9155/55/5/R01.
  2. Keshtkar M, Takavar A, Zahmatkesh MH, Nedaie HA, Vaezzadeh A, Naderi M. Three-dimensional gel dosimetry for dose volume histogram verification in compensator-based IMRT. Int J Radiat Res. 2014;12(1):13-20.
  3. Keshtkar M, Takavar A, Zahmatkesh M, Vaezzadeh A, Gholami M, Ghasemian Z. Application of Polymer Gel dosimetry in Dose Verification of IMRT. Frontiers in Biomedical Technologies. 2015;1(4):279-83.
  4. Watanabe Y, Kubo H. A variable echo-number method for estimating R2 in MRI-based polymer gel dosimetry. Med Phys. 2011;38:975-82. doi.org/10.1118/1.3544659. PubMed PMID: 21452734. PubMed PMCID: 3041809.
  5. De Deene Y, editor. Review of quantitative MRI principles for gel dosimetry. Journal of Physics: Conference Series; 2009: IOP Publishing.
  6. B Baldock C, Murry P, Kron T. Uncertainty analysis in polymer gel dosimetry. Phys Med Biol. 1999;44:N243-6. doi.org/10.1088/0031-9155/44/11/402. PubMed PMID: 10588291.
  7. Baldoc C, Lepage M, Back SA, Murry PJ, Jayasekera PM, Porter D, et al. Dose resolution in radiotherapy polymer gel dosimetry: effect of echo spacing in MRI pulse sequence. Phys Med Biol. 2001;46:449-60. doi.org/10.1088/0031-9155/46/2/312. PubMed PMID: 11229725.
  8. Fong PM, Keil DC, Does MD, Gore JC. Polymer gels for magnetic resonance imaging of radiation dose distributions at normal room atmosphere. Phys Med Biol. 2001;46:3105-13. doi.org/10.1088/0031-9155/46/12/303. PubMed PMID: 11768494.
  9. Watanabe Y, Gopishankar N. Three-dimensional dosimetry of TomoTherapy by MRI-based polymer gel technique. J Appl Clin Med Phys. 2011;12:3273. PubMed PMID: 21330972. PubMed PMCID: 3378327.
  10. Oldham M, Baustert I, Lord C, Smith TA, McJury M, Warrington AP, et al. An investigation into the dosimetry of a nine-field tomotherapy irradiation using BANG-gel dosimetry. Phys Med Biol. 1998;43:1113-32. doi.org/10.1088/0031-9155/43/5/005. PubMed PMID: 9623644.
  11. Gustavsson H, Karlsson A, Back SA, Olsson LE, Haraldsson P, Engstrom P, et al. MAGIC-type polymer gel for three-dimensional dosimetry: intensity-modulated radiation therapy verification. Med Phys. 2003;30:1264-71. doi.org/10.1118/1.1576392. PubMed PMID: 12852552.
  12. Watanabe Y. Variable transformation of calibration equations for radiation dosimetry. Phys Med Biol. 2005;50:1221-34. doi.org/10.1088/0031-9155/50/6/012. PubMed PMID: 15798318.
  13. De Deene Y, Baldock C. Optimization of multiple spin-echo sequences for 3D polymer gel dosimetry. Phys Med Biol. 2002;47:3117-41. doi.org/10.1088/0031-9155/47/17/306. PubMed PMID: 12361214.
  14. De Deene Y, Van de Walle R, Achten E, De Wagter C. Mathematical analysis and experimental investigation of noise in quantitative magnetic resonance imaging applied in polymer gel dosimetry. Signal Processing. 1998;70:85-101. doi.org/10.1016/S0165-1684(98)00115-7.