Document Type : Original Research
Authors
Abstract
Introduction: In radiation therapy centers across Iran, protection of normal tissues is usually accomplished by either Cerrobend or lead block shielding. In this study, the influence of these two shielding methods on central axis dose distribution of photon beam a Cobalt unit was investigated in clinical conditions. Materials and Methods: All measurements were performed for 60Co γ-ray beams and the Cerrobend blocks were fabricated by commercial Cerrobend materials. Standard lead block shields belonged to Cobalt unit. Data was collected through a calibrated ionization chamber, relative dosimetry systemsand a TLD dosimetery.Results: Results of the percent depth dose (PDD) measurements at depths of 0.5, 1, 5, 10, 15 and 20 cm for 23 different field sizes of patients with head and neck cancer showed no significant differences between lead and Cerrobend shielding methods. Measurement results of absolute dosimetry in depths of 1.5, 3, 5, 7, 10 and 12 cm also showed no significant differences between these two shielding methods. The same results were obtained by TLD dosimetry on patient skin.  Conclusion: Use of melt shielding methods is a very easy and fast shield-making technique with no differences in PDD, absolute and skin dose between lead and Cerrobend block shielding methods.
Keywords