Document Type : Original Research

Authors

1 Department of Medical Physics, Babol University of Medical Sciences, Babol, Iran

2 Chronic Respiratory Diseases Research Center, National Research Institute of Tuberculosis and Lung Diseases (NRITLD), Shahid Beheshti University of Medical Sciences, Tehran, Iran

3 PET/CT and Cyclotron Center, Masih Daneshvari Hospital, Shahid Beheshti University of Medical Sciences, Tehran, Iran

4 Department of Medical Physics, Iran University of Medical Sciences, Tehran, Iran

5 Department of Medical Physics and Biomedical Engineering, Tehran University of Medical Sciences, Tehran, Iran

6 Research Center for Molecular and Cellular Imaging, Tehran University of Medical Sciences, Tehran, Iran

Abstract

Background: PET/CT imaging using [18F]-FDG is utilized in clinical oncology for tumor detecting, staging and responding to therapy procedures. Essential consideration must be taken for radiation staff due to high gamma radiation in PET/CT and cyclotron center. The aim of this study was to assess the staff exposure regarding whole body and organ dose and to evaluate environment dose in PET/CT and cyclotron center.
Materials and Methods: 80 patients participated in this study. Thermoluminescence, electronic personal dosimeter and Geiger–Muller dosimeter were also utilized for measurement purpose.
Results: The mean annual equivalent organ dose for scanning operator with regard to lens of eyes, thyroid, breast and finger according to mean±SD value, were 0.262±0.044, 0.256±0.046, 0.257±0.040 and 0.316±0.118, respectively. The maximum and minimum estimated annual whole body doses were observed for injector and the chemist group with values of (3.98±0.021) mSv/yr and (1.64±0.014) mSv/yr, respectively. The observed dose rates were 5.67 µSv/h in uptake room at the distance of 0.5 meter from the patient whereas the value 4.94 and 3.08 µSv/h were recorded close to patient’s head in PET/CT room and 3.5 meter from the reception desk.
Conclusion: In this study, the injector staff and scanning operator received the first high level and second high level of radiation. This study confirmed that low levels of radiation dose were received by all radiation staff during PET/CT procedure using 18F-FDG due to efficient shielding and using trained radiation staff in PET/CT and cyclotron center of Masih Daneshvari hospital.

Keywords

  1. Banghart D. Radiation Protection in Newer Medical Imaging Technologies: PET/CT. Radiation Protection Dosimetry. 2009;135:216-7. doi.org/10.1093/rpd/ncp104.
  2. Antic V, Ciraj-Bjelac O, Stankovic J, Arandjic D, Todorovic N, Lucic S. Radiation exposure to nuclear medicine staff involved in PET/CT practice in Serbia. Radiat Prot Dosimetry. 2014;162:577-85. doi.org/10.1093/rpd/ncu001. PubMed PMID: 24464817.
  3. Leide-Svegborn S. Radiation exposure of patients and personnel from a PET/CT procedure with 18F-FDG. Radiat Prot Dosimetry. 2010;139:208-13. doi.org/10.1093/rpd/ncq026. PubMed PMID: 20167792.
  4. Cao Z, Corley JH, Allison J. 18F protection issues: human and gamma-camera considerations. J Nucl Med Technol. 2003;31:210-5. PubMed PMID: 14657287.
  5. Roberts FO, Gunawardana DH, Pathmaraj K, Wallace A, U PL, Mi T, et al. Radiation dose to PET technologists and strategies to lower occupational exposure. J Nucl Med Technol. 2005;33:44-7. PubMed PMID: 15731021.
  6. Leide-Svegborn S. External radiation exposure of personnel in nuclear medicine from 18F, 99mTC and 131I with special reference to fingers, eyes and thyroid. Radiat Prot Dosimetry. 2012;149:196-206. doi.org/10.1093/rpd/ncr213. PubMed PMID: 21571739.
  7. Demir M, Demir B, Sayman H, Sager S, Sabbir Ahmed A, Uslu I. Radiation protection for accompanying person and radiation workers in PET/CT. Radiat Prot Dosimetry. 2011;147:528-32. doi.org/10.1093/rpd/ncq497. PubMed PMID: 21183551.
  8. Demir M, Demir B, Yaşar D, Sayman HB, Halac M, Ahmed AS, et al. Radiation doses to technologists working with 18F-FDG in a PET center with high patient capacity. Nukleonika. 2010;55:107-12.
  9. Medicine EAoN. Best Practice in Nuclear Medicine, Part 1 A Technologist’s Guide. European Association of Nuclear Medicine; 2006.
  10. Quinn B, Holahan B, Aime J, Humm J, St Germain J, Dauer LT. Measured dose rate constant from oncology patients administered 18F for positron emission tomography. Med Phys. 2012;39:6071-9. doi.org/10.1118/1.4749966. PubMed PMID: 23039646.
  11. Madsen MT, Anderson JA, Halama JR, Kleck J, Simpkin DJ, Votaw JR, et al. AAPM Task Group 108: PET and PET/CT shielding requirements. Med Phys. 2006;33:4-15. doi.org/10.1118/1.2135911. PubMed PMID: 16485403.
  12. Hogg P, Testanera G. Principles and practice of PET/CT: part 1: a technologists guide. 2010.
  13. Protection R. ICRP publication 103. Ann ICRP. 2007;37:2.
  14. Guillet B, Quentin P, Waultier S, Bourrelly M, Pisano P, Mundler O. Technologist radiation exposure in routine clinical practice with 18F-FDG PET. J Nucl Med Technol. 2005;33:175-9. PubMed PMID: 16145226.
  15. Lundberg TM, Gray PJ, Bartlett ML. Measuring and minimizing the radiation dose to nuclear medicine technologists. J Nucl Med Technol. 2002;30:25-30. PubMed PMID: 11948264.
  16. Rohren EM, Turkington TG, Coleman RE. Clinical applications of PET in oncology 1. Radiology. 2004;231:305-32. doi.org/10.1148/radiol.2312021185. PubMed PMID: 15044750.
  17. Pant GS, Sharma SK, Rath GK. Finger doses for staff handling radiopharmaceuticals in nuclear medicine. J Nucl Med Technol. 2006;34:169-73. PubMed PMID: 16951286.
  18. Dalianis K, Malamitsi J, Gogou L, Pagou M, Efthimiadou R, Andreou J, et al. Dosimetric evaluation of the staff working in a PET/CT department. Nuclear instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment. 2006;569:548-50. doi.org/10.1016/j.nima.2006.08.140.
  19. Banghart D, Thomas RH. IAEA Safety Series Report Series: No. 60:‘Radiation Protection in Newer Medical Imaging Techniques: Cardiac CT’IAEA Safety Series Report Series: No. 61:‘Radiation protection in Newer Medical Imaging Techniques: CT Colonography’. Radiation Protection Dosimetry. 2010;138:92-6. doi.org/10.1093/rpd/ncp248.
  20. Kumar S, Pandey AK, Sharma P, Shamim SA, Malhotra A, Kumar R. Instantaneous exposure to nuclear medicine staff involved in PET-CT imaging in developing countries: experience from a tertiary care centre in India. Jpn J Radiol. 2012;30:291-5. doi.org/10.1007/s11604-011-0045-4. PubMed PMID: 22246804.
  21. Al-Haj AN, Lobriguito AM, Arafah A, Parker R. Deriving staff and public doses in a PET/CT facility from measured radiation levels using thermoluminescent dosimetry. Radiat Prot Dosimetry. 2011;144:487-91. doi.org/10.1093/rpd/ncq452. PubMed PMID: 21147788.
  22. Zeff BW, Yester MV. Patient self-attenuation and technologist dose in positron emission tomography. Med Phys. 2005;32:861-5. doi.org/10.1118/1.1869552. PubMed PMID: 15895568.
  23. Carnicer A, Sans-Merce M, Baechler S, Barth I, Donadille L, Ferrari P, et al. Hand exposure in diagnostic nuclear medicine with 18 F-and 99m Tc-labelled radiopharmaceuticals-Results of the ORAMED project. Radiation Measurements. 2011;46:1277-82. doi.org/10.1016/j.radmeas.2011.07.019.
  24. Carnicer A, Ginjaume M, Duch M, Vanhavere F, Merce MS, Baechler S, et al. The use of different types of thermoluminescent dosimeters to measure extremity doses in nuclear medicine. Radiation Measurements. 2011;46:1835-8. doi.org/10.1016/j.radmeas.2011.06.047.
  25. Wrixon AD. New ICRP recommendations. Journal of Radiological Protection. 2008;28:161. doi.org/10.1088/0952-4746/28/2/R02.