Document Type : Original Research
Authors
1 Immunology Research Center, Tabriz University of Medical Sciences, Tabriz, Iran
2 Medical Physics Department, Faculty of Medicine, Tabriz University of Medical Sciences, Tabriz, Iran
3 Department of Oral & Maxillofacial Radiology, Faculty of Dentistry, Tabriz University of Medical Sciences, Tabriz, Iran
4 Vocational School of Health Services, Suleyman Demirel University, Isparta, Turkey
5 Radiology Department, Paramedical Faculty, Tabriz University of Medical Sciences, Tabriz, Iran
Abstract
Background: While the benefits of cone-beam computed tomography (CBCT) are well known in maxillofacial imaging, the use of this modality is not risk-free.
Objective: The aim of this study was to evaluate the exposure doses received by patients during maxillofacial imaging with CBCT.
Methods: Entrance surface dose (ESD) was measured by using thermoluminescent dosimeters (TLDs) attached to the eyes lids, parotid glands and thyroid of 64 patients in two imaging centers (A and B). Phantom dosimetry was performed by a cylindrical poly-methyl methacrylate (PMMA) head-size phantom and an ionization chamber for different exposure parameters. NewTom VGi and Planmeca Promax 3D CBCT scanners were used at centers A and B, respectively.
Results: The mean ESD of the eyes, parotid glands and thyroid were 2.57, 2.33 and 0.28 mGy in center A, 0.35, 2.11 and 0.37 mGy in center B, respectively. ESD of the eyes revealed a significant difference in two centers; in center B, it was 86.4% lower than center A. In the phantom dosimetry, the measured doses of NewTom VGi were 2.63 and 2.08 mGy, respectively by changing field of view (FOV) size from 8×8 cm2 (height × diameter) to 6×6 cm2. For Planmeca Promax 3D, it ranged from 0.98 to 3.24 mGy depending on exposure parameters.
Conclusion: There is a wide range of radiation doses dependent on the units, patients and selected scan parameters. Inappropriate selection of exposure settings, especially FOV size, can seriously increase patient dose.
Keywords
- Heydarheydari S, Haghparast A, Eivazi MT. A Novel Biological Dosimetry Method for Monitoring Occupational Radiation Exposure in Diagnostic and Therapeutic Wards: From Radiation Dosimetry to Biological Effects. J Biomed Phys Eng. 2016;6:21-6. PubMed PMID: 27026951. PubMed PMCID: 4795325.
- Khoshdel-Navi D, Shabestani-Monfared A, Deevband MR, Abdi R, Nabahati M. Local-Reference Patient Dose Evaluation in Conventional Radiography Examinations in Mazandaran, Iran. J Biomed Phys Eng. 2016;6:61-70. PubMed PMID: 27672626. PubMed PMCID: 5022756.
- Janbabanezhad Toori A, Shabestani-Monfared A, Deevband MR, Abdi R, Nabahati M. Dose Assessment in Computed Tomography Examination and Establishment of Local Diagnostic Reference Levels in Mazandaran, Iran. J Biomed Phys Eng. 2015;5:177-84. PubMed PMID: 26688796. PubMed PMCID: 4681462.
- Kiljunen T, Kaasalainen T, Suomalainen A, Kortesniemi M. Dental cone beam CT: A review. Phys Med. 2015;31:844-60. doi.org/10.1016/j.ejmp.2015.09.004. PubMed PMID: 26481816.
- Nemtoi A, Czink C, Haba D, Gahleitner A. Cone beam CT: a current overview of devices. Dentomaxillofac Radiol. 2013;42:20120443. doi.org/10.1259/dmfr.20120443. PubMed PMID: 23818529. PubMed PMCID: 3922261.
- Lavanya R, Babu DB, Waghray S, Chaitanya NC, Mamatha B, Nithika M. A Questionnaire Cross-Sectional Study on Application of CBCT in Dental Postgraduate Students. Pol J Radiol. 2016;81:181-9. doi.org/10.12659/PJR.895688. PubMed PMID: 27158283. PubMed PMCID: 4846182.
- Yang CC. Characterization of Scattered X-Ray Photons in Dental Cone-Beam Computed Tomography. PLoS One. 2016;11:e0149904. doi.org/10.1371/journal.pone.0149904. PubMed PMID: 26950435. PubMed PMCID: 4780712.
- Shahidi S, Bahrampour E, Soltanimehr E, Zamani A, Oshagh M, Moattari M, et al. The accuracy of a designed software for automated localization of craniofacial landmarks on CBCT images. BMC Med Imaging. 2014;14:32.
- Rehani MM. Radiological protection in computed tomography and cone beam computed tomography. Ann ICRP. 2015;44:229-35. doi.org/10.1177/0146645315575872. PubMed PMID: 25816279.
- Nardi C, Borri C, Regini F, Calistri L, Castellani A, Lorini C, et al. Metal and motion artifacts by cone beam computed tomography (CBCT) in dental and maxillofacial study. Radiol Med. 2015;120:618-26. doi.org/10.1007/s11547-015-0496-2. PubMed PMID: 25634792.
- Jaju PP, Jaju SP. Cone-beam computed tomography: Time to move from ALARA to ALADA. Imaging Sci Dent. 2015;45:263-5. doi.org/10.5624/isd.2015.45.4.263. PubMed PMID: 26730375. PubMed PMCID: 4697012.
- Lee C, Lee SS, Kim JE, Symkhampha K, Lee WJ, Huh KH, et al. A dose monitoring system for dental radiography. Imaging Sci Dent. 2016;46:103-8. doi.org/10.5624/isd.2016.46.2.103. PubMed PMID: 27358817. PubMed PMCID: 4925646.
- Cordasco G, Portelli M, Militi A, Nucera R, Lo Giudice A, Gatto E, et al. Low-dose protocol of the spiral CT in orthodontics: comparative evaluation of entrance skin dose with traditional X-ray techniques. Prog Orthod. 2013;14:24. doi.org/10.1186/2196-1042-14-24. PubMed PMID: 24325970. PubMed PMCID: 4384968.
- Araki K, Patil S, Endo A, Okano T. Dose indices in dental cone beam CT and correlation with dose-area product. Dentomaxillofac Radiol. 2013;42:20120362. doi.org/10.1259/dmfr.20120362. PubMed PMID: 23520392. PubMed PMCID: 3635780.
- Signorelli L, Patcas R, Peltomaki T, Schatzle M. Radiation dose of cone-beam computed tomography compared to conventional radiographs in orthodontics. J Orofac Orthop. 2016;77:9-15. doi.org/10.1007/s00056-015-0002-4. PubMed PMID: 26747662.
- Nikneshan S, Aghamiri MR, Moudi E, Bahemmat N, Hadian H. Dosimetry of Three Cone Beam Computerized Tomography Scanners at Different Fields of View in Terms of Various Head and Neck Organs. Iran J Radiol. 2016;13:e34220. doi.org/10.5812/iranjradiol.34220. PubMed PMID: 27853498. PubMed PMCID: 5107245.
- Soares MR, Batista WO, Antonio Pde L, Caldas LV, Maia AF. Study of effective dose of various protocols in equipment cone beam CT. Appl Radiat Isot. 2015;100:21-6. doi.org/10.1016/j.apradiso.2015.01.012. PubMed PMID: 25665897.
- Kadesjo N, Benchimol D, Falahat B, Nasstrom K, Shi XQ. Evaluation of the effective dose of cone beam CT and multislice CT for temporomandibular joint examinations at optimized exposure levels. Dentomaxillofac Radiol. 2015;44:20150041. doi.org/10.1259/dmfr.20150041. PubMed PMID: 26119344. PubMed PMCID: 4628419.
- Chambers D, Bohay R, Kaci L, Barnett R, Battista J. The effective dose of different scanning protocols using the Sirona GALILEOS((R)) comfort CBCT scanner. Dentomaxillofac Radiol. 2015;44:20140287. doi.org/10.1259/dmfr.20140287. PubMed PMID: 25358865. PubMed PMCID: 4614170.
- Batista WO, Soares MR, de Oliveira MV, Maia AF, Caldas LV. Assessment of protocols in cone-beam CT with symmetric and asymmetric beams usingeffective dose and air kerma-area product. Appl Radiat Isot. 2015;100:16-20. doi.org/10.1016/j.apradiso.2015.01.014. PubMed PMID: 25620114.
- Pauwels R, Zhang G, Theodorakou C, Walker A, Bosmans H, Jacobs R, et al. Effective radiation dose and eye lens dose in dental cone beam CT: effect of field of view and angle of rotation. Br J Radiol. 2014;87:20130654. doi.org/10.1259/bjr.20130654. PubMed PMID: 25189417. PubMed PMCID: 4170857.
- Hofmann E, Schmid M, Sedlmair M, Banckwitz R, Hirschfelder U, Lell M. Comparative study of image quality and radiation dose of cone beam and low-dose multislice computed tomography--an in-vitro investigation. Clin Oral Investig. 2014;18:301-11. doi.org/10.1007/s00784-013-0948-9. PubMed PMID: 23460022.
- Rottke D, Patzelt S, Poxleitner P, Schulze D. Effective dose span of ten different cone beam CT devices. Dentomaxillofac Radiol. 2013;42:20120417. doi.org/10.1259/dmfr.20120417. PubMed PMID: 23584925. PubMed PMCID: 3699973.
- Pauwels R, Beinsberger J, Collaert B, Theodorakou C, Rogers J, Walker A, et al. Effective dose range for dental cone beam computed tomography scanners. Eur J Radiol. 2012;81:267-71. doi.org/10.1016/j.ejrad.2010.11.028. PubMed PMID: 21196094.
- Pauwels R, Cockmartin L, Ivanauskaite D, Urboniene A, Gavala S, Donta C, et al. Estimating cancer risk from dental cone-beam CT exposures based on skin dosimetry. Phys Med Biol. 2014;59:3877-91. doi.org/10.1088/0031-9155/59/14/3877. PubMed PMID: 24957710.
- Cassola VF, Milian FM, Kramer R, de Oliveira Lira CA, Khoury HJ. Standing adult human phantoms based on 10th, 50th and 90th mass and height percentiles of male and female Caucasian populations. Phys Med Biol. 2011;56:3749-72. doi.org/10.1088/0031-9155/56/13/002. PubMed PMID: 21628776.
- Marine PM, Stabin MG, Fernald MJ, Brill AB. Changes in radiation dose with variations in human anatomy: larger and smaller normal-stature adults. J Nucl Med. 2010;51:806-11. doi.org/10.2967/jnumed.109.073007. PubMed PMID: 20395339. PubMed PMCID: 2931266.
- Al Najjar A, Colosi D, Dauer LT, Prins R, Patchell G, Branets I, et al. Comparison of adult and child radiation equivalent doses from 2 dental cone-beam computed tomography units. Am J Orthod Dentofacial Orthop. 2013;143:784-92. doi.org/10.1016/j.ajodo.2013.01.013. PubMed PMID: 23726328.
- Sina S, Zeinali B, Karimipoorfard M, Lotfalizadeh F, Sadeghi M, Zamani E, et al. Investigation of the entrance surface dose and dose to different organs in lumbar spine imaging. J Biomed Phys Eng. 2014;4:119-26. PubMed PMID: 25599058. PubMed PMCID: 4289519.
- Mortazavi SMJ, Shareghi A, Kavousi A, Ghiassi-Nejad M, Jafari-Zadeh M, Nazeri F, et al. The need for national diagnostic reference levels: Entrance surface dose measurement in intraoral radiography. Iran J Radiat Res. 2004;2:127-33.
- Zenobio EG, Zenobio MA, Nogueira MS, Silva TA, Shibli JA. Absorbed radiation doses during tomographic examinations in dental implant planning: a study in humans. Clin Implant Dent Relat Res. 2012;14:366-72. doi.org/10.1111/j.1708-8208.2010.00277.x. PubMed PMID: 20491821.
- Bahreyni Toossi MT, Akbari F, Bayani Roodi S. Radiation exposure to critical organs in panoramic dental examination. Acta Med Iran. 2012;50:809-13. PubMed PMID: 23456522.
- Akyalcin S, English JD, Abramovitch KM, Rong XJ. Measurement of skin dose from cone-beam computed tomography imaging. Head Face Med. 2013;9:28. doi.org/10.1186/1746-160X-9-28. PubMed PMID: 24192155. PubMed PMCID: 3851731.
- Goren AD, Prins RD, Dauer LT, Quinn B, Al-Najjar A, Faber RD, et al. Effect of leaded glasses and thyroid shielding on cone beam CT radiation dose in an adult female phantom. Dentomaxillofac Radiol. 2013;42:20120260. doi.org/10.1259/dmfr.20120260. PubMed PMID: 23412460. PubMed PMCID: 3667524.
- Qu XM, Li G, Ludlow JB, Zhang ZY, Ma XC. Effective radiation dose of ProMax 3D cone-beam computerized tomography scanner with different dental protocols. Oral Surg Oral Med Oral Pathol Oral Radiol Endod. 2010;110:770-6. doi.org/10.1016/j.tripleo.2010.06.013. PubMed PMID: 20952220.